• UNIT 12: MEDICAL IMAGING


    Key unit Competence
    By the end of the unit the learner should be able to analyze the processes in medical imaging.
     
    Learning objectives:

      • Outline specific purposes of imaging techniques
      • Explain the effects of various imaging techniques for particular purposes.
      • Explain the basic functioning principles of major medical imaging techniques
      • Identify advantages and disadvantages of medical imaging techniques

    Introductory activity: Investigation on the use of medical imaging techniques:

           
                        fig.12. 1: Medical imaging techniques

    Years ago, the only way to get information from inside of human bodies was through surgery. In modern medicine, medical imaging has undergone major advancements and this ability to obtain information about different parts of the human body has many useful clinical applications.

    Using information provided on the above pictures, answer to the following questions: 
       1. Observe the image A, B, and C (Fig.12.1) and describe what is happening. 
       2. Suggest the technique that is being used for each image? 
       3. Explain the working principle of the mentioned techniques in question 2?

    12.1 X-RAY IMAGING. 

    12.1.1 Interaction of X-rays with matter.

     a. Introduction In unit 


    10, we learnt that X-rays are electromagnetic radiation produced by focusing a beam of high energy electron on a target material in x-ray tube.  Since the major part of the energy of the electrons is converted into heat in the target (only about 1% will appear as X-rays), the target material should have a high melting point and good heat conduction ability. To get a high relative amount of X-ray energy, the anode material should be of high atomic number. Tungsten is the dominating anode material and is in modern X-ray tubes often mixed with Rhenium. 

    In X-ray diagnostics, radiation that is partly transmitted through and partly absorbed in the irradiated object is utilized. An X-ray image shows the variations in transmission caused by structures in the object of varying thickness, density or atomic composition. 
              

    Activity 12.1: Exploration of x-ray image 
    Analyze the figure below answer to the following questions:
                   
                 Fig.12.3 procedure of X-ray image formation
     1. What do the letters A, B, C, D and E stand for?
     2. Explain the process used from A to E.
     3. Explain how radiology differ from radiography.

    b. Attenuation and Absorption of X-rays 

    There are principally two interaction processes that give rise to the x-ray attenuation (variation in photon transmission) through the patient which is the basis of X-ray imaging. These are photoelectric absorption and scattering processes.
             
    A photon which has experienced an interaction process has either been absorbed or has changed its energy and/or direction of motion. A photon that changes its direction of motion is called a scattered photon. For monoenergetic x-ray photons, the number of photons that experience such interactions and therefore removed from the primary beam when this is incident on a thin layer of material is proportional to the number of incident photons (N) and the thickness of the layer (dx) following the expression : 
                                          
    where µ is a constant of proportionality called the linear attenuation coefficient. Integrating the above equation will result in 
                                   
    where  is the initial number of photons in the incident beam. It can be seen that the incident beam photons (or the beam energy) is attenuated exponentially as the X-rays travel through the material. The different interaction processes involved, that are absorption, coherent and incoherent scattering and pair production, add their contributions to the total linear attenuation coefficient

    C. Contrast 

    The contrast is a measure of the difference in radiation transmission or other parameters between two adjacent areas in a radiographic image.  Contrast plays an important role in the ability of a radiologist to perceive image detail.
                                                       
    where ε1and ε2 are energies of the monoenergetic X-ray photons per unit area reaching the detector and therefore absorbed in the detector without and with the contrasting detail respectively. In the case where the film is used as image receptor, the signal is obtained in terms of the optical density. The image contrast is then usually defined as the optical density difference beside and behind the contrasting detail.
    In such situation where monoenergetic photons are considered and no scattered radiation is reaching the detector, the absorbed energy in the detector can be written as

    The contrast is then proportional to the difference in the linear attenuation coefficients and the thickness of the contrasting detail. Therefore, when scattered radiation is neglected in the process, the contrast is independent of the thickness d of the object but also it does not depend on where in the object the contrasting detail is situated. 
    The ability of conventional radiography to display a range of organs and structures may be enhanced by the use of various contrast materials, also known as contrast media. The most common contrast materials are based on barium or iodine. Barium and iodine are high atomic number materials that strongly absorb X-rays and are therefore seen as dense white on radiography. 

    12.1.2   X-rays ImagingTechniques 

    a. Conventional Radiography 

    X-rays are able to pass through the human body and produce an image of internal structures. The resulting image is called a radiograph, more commonly known as an ‘X-ray’ or ‘plain film’. The common terms ‘chest X-ray’ and ‘abdomen X-ray’ are widely accepted and abbreviated to CXR and AXR.
    As a beam of X-rays passes through the human body, some of the X-rays photons are absorbed or scattered producing reduction or attenuation of the beam with the internal human structure acting as contrasting details. 
    Therefore tissues of high density and/or high atomic number cause more X-ray beam attenuation and are shown as lighter grey or white on radiographs. Less dense tissues and structures cause less attenuation of the X-ray beam, and appear darker on radiographs than tissues of higher density. The figure below shows the typical conventional radiograph of a human body.
                            
    Fig.12.6: The five principal radiographic densities. This radiograph of a benign lipoma (arrows) in a child’s thigh demonstrates the five basic radiographic densities: (1) air; (2) fat; (3) soft tissue; (4) bone; (5) metal. (David A Disle (2012) Imaging for students. Fourth Edition. (Page 1)) 
    Five principal densities are easily recognized on this plain radiograph due to the increase in their densities: 
      1. Air/gas appears as black, e.g. lungs, bowel and stomach 
      2.  Fat is shown by dark grey, e.g. subcutaneous tissue layer, retroperitoneal fat 
      3. Soft tissues/water appears as light grey, e.g. solid organs, heart, blood vessels, muscle and fluid-filled organs such as bladder
      4.  Bone appears as off-white
      5. Contrast material/metal: bright white.

    In the past, X-ray films were processed in a darkroom or in freestanding daylight processors. In modern practice, radiographic images are produced digitally using one of two processes, computed radiography (CR) and digital radiography (DR). 
    DR uses a detector screen containing silicon detectors that produce an electrical signal when exposed to X-rays. This signal is analyzed to produce a digital image. Digital images obtained by CR and DR are sent to viewing workstations for interpretation. Images may also be recorded on X-ray film for portability and remote viewing.
                  
    The image given by a computer radiography may be reviewed and reported on a computer workstation. This allows various manipulations of images as well as application of functions such as measurements of length and angles measurements. 
     The relative variance of the shadows depends upon the density of the materials within the object or body part. Dense, calcium – rich bone absorbs X-rays to a higher degree than soft tissues that permit more X-rays to pass through them, making X-rays very useful for capturing images of bone.
    In projection radiography, there is much room for adjusting the energy level of the X-rays depending on the relative densities of the tissues being imaged and also how deep through a body the waves must travel in order to achieve the imaging.

     • Images of bones (for instance, to examine a fracture or for diagnostic measures related to bone conditions like osteoarthritis or certain cancers) require highenergy X-rays because of the high density of bone.
     • Images of soft tissues like lungs, heart and breasts (both chest X-rays and mammography are very common diagnostic applications of X-rays) require relatively less energy from the X-rays in order to penetrate properly and achieve excellent images.
     • In order to achieve these different energies, technologists use X-ray generators of different voltages and equipped with anodes made of different metals.

    b. Mammography

    Activity 12.2 

    One day a girl suffering from the breast tells her mother about the problem. And her mother advises her to go to the hospital to consult a doctor.
            a. Think of the problem that girl may have.
            b. Try to explain what may be the cause of that problem.
            c. If you are a doctor how can you detect such problem?
            d. Which advise can you give to other people who are not suffering from that problem in order to prevent it?

    Mammography is a specialized medical imaging that uses low-dose X-rays to investigate the internal structure of the breast. A mammography exam, called a mammogram, helps in the early detection and diagnosis of women’s breast diseases such as breast cancer before even experiencing any symptom. Below is a typical mammography test showing the presence of abnormal areas of density, mass, or calcification that may indicate the presence of cancer.       

    Mammographysada) the breast is pressed between two plates x-rays are used to takes pictures of breast tissues,(b)photographic image of breast tissues, (c)Breast with cancer. 
    A mammography unit is a rectangular box that houses the tube in which X-rays are produced. The unit is used exclusively for X-ray exams of the breast, with special accessories that allow only the breast to be exposed to the X-rays. Attached to the unit is a device that holds and compresses the breast and positions it so images can be obtained at different angles.

    In conventional film and digital mammography, a stationery X-ray tube captures an image from the side and an image from above the compressed breast. Breast tomosynthesis, also called three-dimensional (3-D) mammography and digital breast tomosynthesis (DBT), is an advanced form of breast imaging where multiple images of the breast from different angles are captured and reconstructed (“synthesized”) into a three-dimensional image set. In this way, 3-D breast imaging is similar to computed tomography (CT) imaging in which a series of thin “slices” are assembled together to create a 3-D reconstruction of the body.

    c. Computer tomography scan (ct scan) 

    a. CT terminology 

    In 1970s, a revolutionary new X-ray technique was developed called Computer tomography (CT), which produce an image of a slice through the body. The word tomography comes from the Greek: tomos =slice, graph= picture.) 

    A computed tomography scan also known as CT scan, makes use of computerprocessed combinations of many X-ray measurements taken from different angles to produce cross-sectional (tomographic) images (virtual “slices”) of specific areas of a scanned object, allowing the user to see inside the object without cutting it. Other terms include computed axial tomography (CAT scan) and computer aided tomography. 

    The term “computed tomography” (CT) is often used to refer to X-ray CT, because it is the most commonly known form but many other types of CT exist. 

    CT is an imaging technique whereby cross-sectional images are obtained with the use of X-rays. In CT scanning, the patient is passed through a rotating gantry that has an X-ray tube on one side and a set of detectors on the other. Information from the detectors is analysed by computer and displayed as a grey-scale image. Owing to the use of computer analysis, a much greater array of densities can be displayed than on conventional X-ray films.

    This allows accurate display of cross-sectional anatomy, differentiation of organs and pathology, and sensitivity to the presence of specific materials such as fat or calcium. As with plain radiography, high- density objects cause more attenuation of the X-ray beam and are therefore displayed as lighter grey than objects of lower density. 

    b. Principle behind of computer tomography scan (CT scan). Computer Tomography is shown in below figure: a thin collimated beam of X- ray(“ to collimate” means to “make straight”) passes through the body to a detector that measures the transmitted intensity. Measurements are made at a large number of points as the source and detector are moved past the body together. The apparatus is rotated slightly about the body axis and again scanned; this is repeated at 1st intervals for  180 . The intensity of the transmitted beam for the many points of each scan, and for each angles, are sent to a computer that reconstructs the image of the slice. Note that the imaged slice is perpendicular to the long axis of the body. For this reason, CT is sometimes called computerize axial tomography.

    The use of single detector would require a few minutes for many scans needed to form a complete image. Much faster scanner use a fan beam in which passing through the entire cross section of the body are detected simultaneously by many detectors.The x-ray source and the detectors are rotated about the patient and an image requires only few seconds to be seen. This means that rays transmitted through the entire body are measured simultaneously at each angle where the source and detector rotate to take measurements at different angles.


    CT images of internal organs, bones, soft tissue, and blood vessels provide greater clarity and more details than conventional X-ray images, such as a chest X-Ray.

    Function of CT scan
     • A motorized table moves the patient through a circular opening in the CT imaging system.
     • While the patient is inside the opening, an X-ray source and a detector assembly within the system rotate around the patient. A single rotation typically takes a second or less. During rotation the X-ray source produces a narrow, fan-shaped beam of X-rays that passes through a section of the patient’s body.
     • Detectors in rows opposite the X-ray source register the X-rays that pass through the patient’s body as a snapshot in the process of creating an image. Many different “snapshots” (at many angles through the patient) are collected during one complete rotation.
     • For each rotation of the X-ray source and detector assembly, the image data are sent to a computer to reconstruct all of the individual “snapshots” into one or multiple cross-sectional images (slices) of the internal organs and tissues.

    Note that, it is advisable to avoid unnecessary radiation exposure; a medically needed CT scan obtained with appropriate acquisition parameter has benefits that outweigh the radiation risks. 

    12.1.3 Checking my progress

       1. Outline the advantage and disadvantages CT scan
       2.  Explain the types of x-ray imaging used in mammography.
       3. In mammography exams, is the breast compression necessary?
           Why
       4. A beam of X-rays passes through the human body of tissues with different        densities; explain the production of X-rays on less dense tissues
    12.2 ULTRASONIC IMAGING
     

    12.2.1 Basics of Ultra sound and its production 

    Activity 12.3
     
     1. Distinguish ultrasound from infrasound?
     2. Where do you think ultrasound may be applied in daily life?
     3. Advise on how ultrasound be used in medicine?

    Sound can refer to either an auditory sensation in the brain or the disturbance in a medium that causes this sensation. Hearing is the process by which the ear transforms sound vibrations into nerve impulses that are delivered to the brain and interpreted as sounds. Sound waves are produced when vibrating objects produce pressure pulses of vibrating air. The auditory system can distinguish different subjective aspects of a sound, such as its loudness and pitch.

    Pitch is the subjective perception of the frequency, which in turn is measured in cycles per second, or Hertz (Hz). The normal human audible range extends from about 20 Hz to 20 000 Hz, but the human ear is most sensitive to frequencies of 1 000 Hz to 4 000 Hz. Loudness is the perception of the intensity of sound, related to the pressure produced by sound waves on the tympanic membrane. The pressure level of sound is measured in decibels (dB), a unit for comparing the intensity of a given sound with a sound that is just perceptible to the normal human ear at a frequency in the range to which the ear is most sensitive. On the decibel scale, the range of human hearing extends from 0 dB, which represents the auditory threshold, to about 130 dB, the level at which sound becomes painful.

    12.2.2 Interaction of sound waves with different structure inside the body 
    a. Introduction 

    Ultrasound imaging uses ultra-high-frequency sound waves to produce crosssectional images of the body. Ultrasound is actually sound with a frequency in excess of 20 kHz, which is the upper limit of human hearing. Typical ultrasound frequencies used for clinical purposes are in the 2 MHz to 10 MHz range.

    Different tissues in a human or animal body alter the ultra sound waves in different ways. Some waves are reflected directly while others scatter the waves before they return to the transducer as echoes. The reflected ultrasound pulses detected by the transducer need to be amplified in the scanner or ultrasonic probe. The echoes that come from deep within the body are more attenuated than those from the more superficial parts and therefore required more amplification. 
     When echoes return to the transducer, it is possible to reconstruct a two dimensional map of all the tissues that have been in the beams. The information is stored in a computer and displayed on a video (television) monitor. Strong echoes are said to be of the high intensity and appear as brighter dots on the screen.

    b. Reflection of ultrasound 

    When the pulse of ultrasound is sent into the body and meets a boundary between two media, of different specific acoustic impedance Z, the sound wave needs to change gear in order to continue. If the difference in Z across the boundary is large 
    the wave cannot easily adjust: there is an “acoustic mismatch”. Most of the wave is reflected and a strong echo is recorded. The fraction of intensity reflected back  to that incident  at the normal incidence, is known as the intensity of reflection coefficient α 
                  

    Ultrasounds are high-frequency sound waves above the human ear’s audible range: that is with frequency sound waves greater than 20 kHz. In fact, the frequencies used in medicine are much higher than this, typically between 1 MHz and 15 MHz. like all sound waves, ultrasound consists of longitudinal, elastic or pressure waves, capable of traveling through solids, liquids and gases. This makes them ideal for penetrating the body, unlike transverse mechanical waves, which cannot travel to any great extent through fluids.

    c. Attenuation of ultrasound 

    The attenuation of the waves describes the reduction in its intensity as they travel through a medium. This loss is due to a number of factors:
      • The wave simply “spreads out” and suffers an “inverse square law type” reduction in intensity.
      • The wave is scattered away from its original direction
      • The wave is absorbed in the medium.

    The amount of absorption of ultrasound beam in a medium is described by the absorption coefficient α , which is intensity level per unit length. It is expressed in decibels per cm and it firstly depends on the type of medium the wave is propagating into. As example whilst water absorbs very little ultrasound, bone is a strong absorber, putting it at risk, for example, during high- power ultrasound therapy. 

    Secondly, higher frequencies suffer greater absorption. In fact if the frequency is doubled, the absorption increases by the factor of four. This has very important consequences when choosing the best frequency at which to image the body. If the selected frequency is too high, the ultrasound will not be able to penetrate to the regions under investigation.  

    12.2.3 Ultrasonic imaging techniques 

    The basic component of the ultrasound probe is the piezoelectric crystal. Excitation of this crystal by electrical signals causes it to emit ultra-high-frequency sound waves; this is the piezoelectric effect. The emitted ultrasound waves are reflected back to the crystal by the various tissues of the body. These reflected sound waves also called the “echoes” act on the piezoelectric crystal in the ultrasound probe to produce an electric signal, again by the piezoelectric effect. It is this electric signal which is analysed by a computer produces a cross-sectional image.
                               
    The process of imaging is the same as the echo-locating sonar of a submarine or a bat. The observer sends out a brief pulse of ultrasound and waits for an echo. The pulse travels out, reflects off the target and returns. The ultrasound machine uses pulses because the same device acts as both transmitter and receiver. If it continually sent out sounds, then the receiver would not hear the much softer echo over the louder transmission.

    The duty cycle of the ultrasound imager is the amount of time spent transmitting compared to the total time of transmitting and listening. Sonar is an acronym for Sound Navigation and Ranging. It relies on the reflection of ultrasound pulses. A short pulse of ultrasound is directed towards the object interest, which then reflects it back as an echo. The total time  between transmission of pulse and reception of an echo is measured, often using a cathode ray oscilloscope (CRO). The sonar principle is used to estimate the depth of a structure, using
                       
    Where tis the time taken to go and back and v is the velocity of ultrasound in the medium. The factor of 2 is necessary because the pulse must travel “there and back” An ultrasound beam structure is directly into the body. The reflection or echoes from different body structure are then detected and analyzed, yielding information about the locations. For example if the time delays between the reception of echo pulse1 and 2 (Fig.12.7 below) is t, then the diameter of the baby’s head can be found using the above formula.

                     
    During an investigation using ultrasound, the time delay for an echo to return from a structureis 10.5 s µ If the average velocity of ultrasound in the eye is 1510 m/s. Calculate the depth of the structure. 
             
    a. Doppler ultrasonic 

    An object travelling towards the listener causes sound waves to be compressed giving a higher frequency; an object travelling away from the listener gives a lower frequency. The Doppler effect has been applied to ultrasound imaging. Flowing blood causes an alteration to the frequency of sound waves returning to the ultra sound probe. This frequency change or shift is calculated allowing quantization of blood flow. The combination of conventional two-dimensional ultra sound imaging with Doppler ultra sound is known as Duplex ultra sound.

                         
    Fig.12.11: Duplex ultra sound. The Doppler sample gate is positioned in the artery (arrow) and the frequency shifts displayed as a graph. Peak systolic and end diastolic velocities are calculated and also displayed on the image in centimeters per second.
    As ultrasound imaging uses sound waves to produce pictures of inside of the body. It is used to help diagnose the cause of pain, swelling and infection in the body’s internal organs and to examine a baby in pregnant woman and the brain and hips in infants. It is also used to help guide biopsies, diagnose heart conditions and assess damage after a heart attack. 
    Ultrasound examinations do not use ionizing radiation (x-rays), there is no radiation exposure to the patient. Because ultrasound image are captured in real time, they can show the structure and movement of the body’s internal organs, as well as blood flowing through blood vessels.

    b. Advantages and Disadvantages of ultrasounds 

    The advantages of ultrasound over other imaging modalities include:
      • Lack of ionizing radiation.
      • Relatively low cost
      • It is noninvasive (of medicine procedures not involving the introduction of instruments into the body)
      • Quick procedure
      • Good for examining soft tissues.
      • Portability of equipment.

    Some disadvantages of ultrasound include: 

     • It is highly operator dependent as it relies on the operator to produce and interpret images at the time of examination
     • Not as much details as X-rays and MRI
     • It cannot be used in areas that contain gas (such as lungs)
     • Doesn’t pass through bones.
     • Can be wrong in detecting physical abnormalities. 

    12.2.4 Checking my progress 

     1. Explain how ultrasound imaging is used?
     2. Who take the decision to scan or not to scan in normal pregnancy?
     3. What are the risks and side effects to the mother or baby during ultrasound? 4. If an ultrasound is done at 6 to 7 weeks and a heartbeat is not detected, does

    12.3 SCINTIGRAPHY (NUCLEAR MEDICINE) 

    Activity 12.4

      a. What do you understandby ‘radionuclide imaging’?
      b. What is a radionuclide scan used for?
      c. Compare radionuclide scan with mammography scan?

    12.3.1 Physics of scintigraphy and terminology
     

    Scintigraphy refers to the use of gamma radiation to form images following the injection of various radiopharmaceuticals. The key word to understanding of scintigraphy is radiopharmaceutical. ‘Radio’ refers to the radionuclide, i.e. the emitter of gamma rays.
     The most commonly used radionuclide in clinical practice is technetium, written in this text as mTc 99 , where 99 is the atomic mass, and the ‘m’ stands for metastable. Metastable means that the technetium atom has two basic energy states: high and low energy states. As the technetium transforms from the high-energy state to the low-energy state, it emits a quantum of energy in the form of a gamma ray, which has energy of 140 keV. Other commonly used radionuclides include gallium citrate


    12.3.2 Basic functioning of radionuclide scan 

    Activity12.6
     
    For radionuclide imaging, it is advisable for the patient to consume a small quantity of radionuclide, or it is injected into a vein in your arm. a. How long does it take? b. What is the purpose of those radionuclide chemicals? c. Assuming the patient has already consumed the radionuclide for him/ her to be scanned and wants to take another scan on another part of the body, will the patient be required to take another dose of the nuclide? d.  You as a student, what advice can you give to a patient who develops allergies after taking the radio nuclear chemical?

    A radionuclide scan is a way of imaging bones, organs and other parts of the body by using a small dose of a radioactive chemical. There are different types of radionuclide chemical. The one used depends on which organ or part of the body need to be scanned.

    A radionuclide (sometimes called a radioisotope or isotope) is a chemical which emits a type of radioactivity called gamma rays. A tiny amount of radionuclide is put into the body, usually by an injection into a vein. Sometimes it is breathed in, or swallowed, or given as eye drops, depending on the test. 

    Gamma rays are similar to X-rays and are detected by a device called a gamma camera. The gamma rays which are emitted from inside the body are detected by the gamma camera, are converted into an electrical signal and sent to a computer. The computer builds a picture by converting the differing intensities of radioactivity emitted into different colors or shades of grey. 

    However, radionuclide imaging techniques do not depict structural anatomy like ultrasound, X-ray computed tomography (XCT) or conventional radiographs. It is the only established noninvasive technique available to investigate organ physiology, although recently Nuclear magnetic resonance (NMR) imaging technique has shown its capability to probe organ physiology and anatomy without ionizing radiation. 

    Radionuclide scans do not generally cause any after effects. Through the natural process of radioactive decay, the small amount of radioactive chemical in your body will lose its radioactivity over time. Although the levels of radiation used in the scan are small, patients may be advised to observe special precautions. 

    12.3.3 Limitations and disadvantages of scintigraphy 

    The main advantages of scintigraphy are its high sensitivity and the fact that the functional information is provided as well as anatomical information. However it has some disadvantages that are listed below:
      I. Generally poor resolution compared with other imaging techniques.
      II. Radiation risks due to the administered radionuclide
     III. Can be invasive, sometimes requiring an injection into the bloodstream 
     IV. Disposal for radioactive waste, including that from patients, requires special procedures.
     V. Relatively high costs associated with radiotracer production and administration.

    12.3.4 Checking my progress 

    Choose the correct answers
     1. Scintigraphy refers to the use of: 
         a.  Gamma radiation to form images
         b. X- ray radiation to form images
         c. X- rays and gamma radiations to form images
         d. None of radiation to form images.
      2. The radionuclide in clinical practice are
          a.   Technetium
          b. Thallium
          c. Gallium 
          d. ALL of them

    12.4 MAGNETIC RESONANCE IMAGING (MRI) 

    Activity 12.7: Principles of MRI

       I. What does MRI mean?
      II. What is it used for?
     III. What makes MRI to be powerful compared to other imaging techniques?
     IV. is it advisable for a pregnant woman to be placed in MRI Scanner?
          Explain your view?

    12.4.1. MRI physics and terminology. 

    Magnetic resonance (MR) imaging has become the dominant clinical imaging modality with widespread, primarily noninvasive, applicability throughout the body and across many disease processes. The progress of MR imaging has been rapid compared with other imaging technologies and it can be attributed in part to physics and in part to the timing of the development of MR imaging, which corresponded to an important period of advances in computing technology. 

    Initially let us described how magnetic resonance can be demonstrated with a pair of magnets and a compass. If a compass happens to find itself near a powerful magnet, the compass needle will align with the field. In a normal pocket compass, the needle is embedded in liquid to dampen its oscillations. Without liquid, the needle will vibrate through the north direction for a period before coming to rest. The frequency of the oscillations depends on the magnetic field and of the strength of the magnetic needle.

                        
    Let us focus on what made the needle oscillate. It was the small movements of the magnet, back and forth, or more precisely the oscillation of a weak magnetic field perpendicular to the powerful stationary magnetic field caused by the movement of the magnet. But oscillating magnetic field is what we understand by “radio waves”, which means that in reality, we could replace the weak magnet with other types of radio wave emitters. 

    This could, for example, be a small coil subject to an alternating current, as shown in figure above. Such a coil will create a magnetic field perpendicular to the magnetic needle. The field changes direction in synchrony with the oscillation of the alternating current, so if the frequency of the current is adjusted to the resonance frequency of the magnetic needle, the current will set the needle in motion. This is also applied in an MR scanner. In summary, the needle can be set in motion from a distance by either waving a magnet or by applying an alternating current to a coil. In both situations, magnetic resonance is achieved when the magnetic field that motion or alternating currents produce, oscillates at the resonance frequency. When the waving or the alternating current is stopped, the radio waves that are subsequently produced by the oscillating needle will induce a voltage over the coil.

    MRI uses the magnetic properties of spinning hydrogen atoms to produce images. The first step in MRI is the application of a strong, external magnetic field. For this purpose, the patient is placed within a large powerful magnet. Most current medical MRI machines have field strengths of 1.5 or 3.0 Tesla. The hydrogen atoms within the patient align in a direction either parallel or antiparallel to the strong external field. 

    A greater proportion aligns in the parallel direction so that the net vector of their alignment, and therefore the net magnetic vector, will be in the direction of the external field. This is known as longitudinal magnetization. A second magnetic field is applied at right angles to the original external field. This second magnetic field is known as the radiofrequency pulse (RF pulse), because it is applied at a frequency in the same part of the electromagnetic spectrum as radio waves. A magnetic coil, known as the RF coil, applies the RF pulse.

    The RF pulse causes the net magnetization vector of the hydrogen atoms to turn towards the transverse plane, i.e. a plane at right angles to the direction of the original, strong external field. The component of the net magnetization vector in the transverse plane induces an electrical current in the RF coil. This current is known as the MR signal and is the basis for formation of an image. Computer analysis of the complex MR signal from the RF receiver coils is used to produce an MR image.

    12.4.2. The magnetism of the body 

    Let‘s see how magnet needles with and without spin are affected by radio waves, we now turn to the “compass needles” in our own bodies.
        a. Most frequently, the MR signal is derived from hydrogen nuclei (meaning the atomic nuclei in the hydrogen atoms). Most of the body’s hydrogen is found in the water molecules. Few other nuclei are used for MR.
        b.  Hydrogen nuclei (also called protons) behave as small compass needles that align themselves parallel to the field.
        c. The compass needles (the spins) are aligned in the field, but due to movements and nuclear interactions in the soup, the alignment only happens partially.
        d. The nuclei in the body move among each other (thermal motion) and the net magnetization in equilibrium is thus temperature dependent. 
        e. Due to the number of hydrogen nuclei (about 27 10 ) found in the body, the net magnetization still becomes measurable. It is proportional to the field: A large field produces a high degree of alignment and thus a large magnetization and better signal to noise ratio.

    12.4.3.  Magnetic ResonanceImaging (MRI). 

    In MRI, a particular type of nucleus is selected and its distribution in the body is monitored. Hydrogen is the most commonly imaged element, not only due to its abundance in the body but also because it gives the strongest MRI signals.

    The technique uses a very powerful magnet to align the nuclei of atoms inside the body, and a variable magnetic field that causes the atoms to resonate, a phenomenon called nuclear magnetic resonance. The nuclei produce their own rotating magnetic fields that a scanner detects and uses to create an image. 

    MRI is used to diagnose a variety of disorders, such as strokes, tumors, aneurysms, spinal cord injuries, multiple sclerosis and eye or inner ear problems. It is also widely used in research to measure brain structure and function, among other things.

    An MRI scan can be used to examine almost any part of the body, including the:   • Brain and spinal cord 
      • Bones and joints
      • Breasts
      • Heart and blood vessels
      • Internal organs, such as the liver, womb or prostate gland ,etc
    The results of an MRI scan can be used to help diagnose conditions, plan treatments and assess how effective previous treatment has been.

    12.4.4. Functional of MRI Scan

     Activity:12.8

       a. Explain the function of MRI Scan.
       b. What are the advantages and disadvantages of MRI Scan?
       c. What are the hazards associated with MRI?

    There are many forms of MRI, some of them are:
      a. Diffusion-weighted imaging. 

    Diffusion-weighted imaging (DWI) is sensitive to the random Brownian motion (diffusion) of water molecules within tissue. The greater the amount of diffusion, the greater the signal loss on DWI. Areas of reduced water molecule diffusion show on DWI as relatively high signal. Diffusion-weighted imaging is the most sensitive imaging test available for the diagnosis of acute cerebral infarction. With the onset of acute ischaemia and cell death there is increased intracellular water (cytotoxicoedema) with restricted diffusion of water molecules. An acute infarct therefore shows on DWI as an area of relatively high signal.

    b. Perfusion-weighted imaging

    In perfusion-weighted imaging (PWI) the brain is rapidly scanned following injection of a bolus of contrast material (gadolinium). The data obtained may be represented in a number of ways including maps of regional cerebral blood volume, cerebral blood flow, and mean transit time of the contrast bolus. PWI may be used in patients with cerebral infarct to map out areas of brain at risk of ischaemia that may be salvageable with thrombolysis.

    c. Magnetic resonance spectroscopy 

    Magnetic resonance spectroscopy (MRS) uses different frequencies to identify certain molecules in a selected volume of tissue, known as a voxel. Following data analysis, a spectrographic graph of certain metabolites is drawn. Metabolites of interest include lipid, lactate, NAA (N-acetylaspartate), choline, creatinine, citrate and myoinositol.

    Uses of MRS include characterization of metabolic brain disorders in children, imaging of dementias, differentiation of recurrent cerebral tumour from radiation necrosis, and diagnosis of prostatic carcinoma. 

    d. Blood oxygen level-dependent imaging 

    Blood oxygen level-dependent (BOLD) imaging is a non-invasive functional MRI (fMRI) technique used for localizing regional brain signal intensity changes in response to task performance. BOLD imaging depends on regional changes in concentration of deoxyhemoglobin, and is therefore a tool to investigate regional cerebral physiology in response to a variety of stimuli. BOLD fMRI may be used prior to surgery for brain tumor or arteriovenous malformation (AVM), as a prognostic indicator of the degree of postsurgical deficit.

    12.4.5 Advantage and disadvantages of MRI. 

    Advantages of MRI in clinical practice include: 
     1. Excellent soft tissue contrast and characterization
     2. Lack of ionizing radiation.
     3. Noninvasive machine.
     4. Lack of artefact from adjacent bones, e.g. pituitary fossa

    Disadvantages of MRI:
     
    1. High capital and running costs.
     2. Image selected and interpretation is complex.
     3. Examination can be difficult for some people who are claustrophobic
     4. The examination is noisy and takes long.
     5. Hazards with implants, particularly pacemakers.
     6. Practical problems associated with large superconducting magnets. 

    12.4.7. Checking my progress 

     1. What is meant by relaxation in the context of MRI?
     2. Give the reasons why the hydrogen nucleus is most used in MRI.
     3. What does NMR stand for? Explain carefully the role of the three terms involved
     4. Draw the basic steps in the formation of MRI image.

    12.5 ENDOSCOPY 

    Activity 12.9 
     
    a. How can we examine inside the stomach by using light rays? 
    2. How is endoscope performed?
     
    12.3.1 Description 

    Endoscopy is a nonsurgical procedure used to examine a person’s digestive tract. Using an endoscope, which is a flexible tube with a light and camera attached to it, the specialist can view pictures of your digestive tract on a monitor.
    Endoscopy is a nonsurgical procedure used to examine a person’s digestive tract. Using an endoscope, which is a flexible tube with a light and camera attached to it, the specialist can view pictures of your digestive tract on a monitor. During an upper endoscopy, an endoscope is easily passed through the mouth and throat and into the esophagus, allowing the specialist to view the esophagus, stomach, and upper part of the small intestine. Similarly, endoscopes can be passed into the large intestine (colon) through the rectum to examine this area of the intestine. 

    12.3.2 Upper endoscopy 

    Upper Endoscopy (also known as gastroscopy, EGD, or esophagogastroduodenoscopy) is a procedure that enables your surgeon to examine the lining of the esophagus (swallowing tube), stomach and duodenum (first portion of the small intestine). A bendable, lighted tube about the thickness of your little finger is placed through your mouth and into the stomach and duodenum.

                      
    12.3.3. How is the upper endoscopy performed? 

    Upper endoscopy is performed to evaluate symptoms of persistent upper abdominal pain, nausea, vomiting, difficulty swallowing or heartburn. It is an excellent method for finding the cause of bleeding from the upper gastrointestinal tract. It can be used to evaluate the esophagus or stomach after major surgery. It is more accurate than X-rays for detecting inflammation, ulcers or tumors of the esophagus, stomach and duodenum. Upper endoscopy can detect early cancer and can distinguish between cancerous and noncancerous conditions by performing biopsies of suspicious areas.
            
    A variety of instruments can be passed through the endoscope that allows the surgeon to treat many abnormalities with little or no discomfort, remove swallowed objects, or treat upper gastrointestinal bleeding. Safe and effective control of bleeding has reduced the need for transfusions and surgery in many patients. 

    12.3.4. Advantages and disadvantages of endoscopy 

    Advantages

     
    • Complete visualization of the entire stomach or digestive tract.
     • It is very safe and effective tool in diagnosis
     • Does not leave any scar because it uses natural body openings. • It is cost effective and has low risk 
     • They are generally painless.
     • Can do therapeutic interventions
     • Allows for sampling/biopsying of small bowel mucosa
     • Allows for resection of polyps.

    Disadvantages: 

    Although the endoscope is very safe; however, the procedure has a few potential complications which may include:
     • Bleeding
     • Perforation (tear in the gut wall)
     • Infection 
     • Reaction to sedation (action of administering a sedative drug to produce a state of calm or sleep.
     • Technically difficult procedure
     • Very time consuming (Procedure can take > 3 hours)
     • Patient may need to be admitted to the hospital
     • Higher risk of small bowel perforation
     • Case reports of pancreatitis and intestinal necrosis
     • Reported incidents of aspiration and pneumonia

    12.3.5 Checking my progress 

       1. What are instruments used to view the esophagus, stomach and upper small intestine of human body?
       2. Explain the function of endoscope.
       3. Compare and contrast colonoscopy and gastroscopy

    12.3.6. hazards associated with medical imaging 

    The following are some hazards associated with medical imaging:
      1. Exposure to ionizing radiation
      2. Anaphylactoid reactions to iodinated contrast media
      3. Contrast-induced nephropathy (CIN)
      4. MRI safety issues
      5. Nephrogenic systemic sclerosis (NSF) due to Gd-containing contrast media.

    1. Exposure to ionizing radiation 

    Radiation effects and effective dose Radiography, scintigraphy and CT use ionizing radiation. Numerous studies have shown that ionizing radiation in large doses is harmful. The risks of harm from medical radiation are low, and are usually expressed as the increased risk of developing cancer as a result of exposure. Radiation effects occur as a result of damage to cells, including cell death and genetic damage. Actively dividing cells, such as are found in the bone marrow, lymph glands and gonads are particularly sensitive to radiation effects.

    2. Anaphylactoid contrast media reactions 

    Most patients injected intravenously with iodinated contrast media experience normal transient phenomena, including a mild warm feeling plus an odd taste in the mouth. With modern iodinated contrast media, vomiting at the time of injection is uncommon. More significant adverse reactions to contrast media may be classified as mild, intermediate or severe anaphylactoid reactions:
     • Mild anaphylactoid reactions: mild urticaria and pruritic
     • Intermediate reactions: more severe urticaria, hypotension and mild bronchospasm
     • Severe reactions: more severe bronchospasm, laryngeal oedema, pulmonary oedema, unconsciousness, convulsions, pulmonary collapse and cardiac arrest.

    3. Contrast-induced nephropathy 


    Contrast-induced nephropathy (CIN) refers to a reduction of renal function (defined as greater than 25 per cent increase in serum creatinine) occurring within 3 days of contrast medium injection. Risk factors for the development of CIN include:  

    Pre-existing impaired renal function, particularly diabetic nephropathy, Dehydration, Sepsis, Age>60 years,  Recent organ transplant ,  Multiple myeloma. The risk of developing CIN may be reduced by the following measures:  Risk factors should be identified by risk assessment questionnaire.
      • Use of other imaging modalities in patients at risk including US or noncontrast-enhanced CT.
      • Use of minimum possible dose where contrast medium injection is required.    • Adequate hydration before and after contrast medium injection.
      • Various pretreatments have been described, such as oral acetylcysteine; however, there is currently no convincing evidence that anything other than hydration is beneficial.

    4. MRI safety issues 

    Potential hazards associated with MRI predominantly relate to the interaction of the magnetic fields with metallic materials and electronic devices. Ferromagnetic materials within the patient could possibly be moved by the magnetic field causing tissue damage. Common potential problems include metal fragments in the eye and various medical devices such as intracerebral aneurysm clips. Patients with a past history of penetrating eye injury are at risk for having metal fragments in the eye and should be screened prior to entering the MRI room with radiographs of the orbits. The presence of electrically active implants, such as cardiac pacemakers, cochlear implants and neurostimulators, is generally a contraindication to MRI unless the safety of an individual device is proven.

    5. Nephrogenic systemic sclerosis 

    Nephrogenic systemic sclerosis (NSF) is a rare complication of some Gd-based contrast media in patients with renal failure. Onset of symptoms may occur from one day to three months following injection. Initial symptoms consist of pain, pruritis and erythema, usually in the legs. As NSF progresses there is thickening of skin and subcutaneous tissues, and fibrosis of internal organs including heart, liver and kidneys. Identifying patients at risk, including patients with known renal disease, diabetes, hypertension and recent organ transplant, may reduce the risk of developing NSF following injection of Gd- based contrast media.

    Risk reduction in MRI 

    A standard questionnaire to be completed by the patient prior to MRI should cover relevant factors such as:
      • Previous surgical history
      • Presence of metal foreign bodies including aneurysm clips, etc.
      • Presence of cochlear implants and cardiac pacemakers 
      • Possible occupational exposure to metal fragments and history of penetrating eye injury 
      • Previous allergic reaction to Gd-based contrast media
      • Known renal disease or other risk factors relevant to NSF.

    END UNIT ASSESSMENT

    Part I: Copy the following in your notebook and chose the correct answer
       1. Which are included in the system components of gamma rays camera for producing image of the body? 
          a.  Collimator b. Scintillation c. Attenuation d.  All of the above

        2. Which of the following modalities does not use a form of ionizing radiation?        a. Radiography. b. Computed tomography. c. Sonography. d. Magnetic resonance imaging
        3. Hazards not associated with modern medical imaging include:
            a. Anaphylactoid reactions to iodinated contrast media 
            b. Complication of some Gd-based contrast media in patients with renal failure. 
            c. Imaging of the breast improves a physician’s ability to detect small tumors 
            d. Radiation effects and effective dose Radiography.

         4. Medical imaging systems are often evaluated the characteristics which are directly related to: 
             a. Image noise.   b. Image blurring.     c. Image unsharpness. d. Visibility of anatomical detail.

         5. Risks associated with radionuclide imaging are: 
               a. Generally poor resolution compared with other imaging modalities.
               b. Rarely receiving an overdose of chemical injected in the vein of the body.  c. High capital and running costs 
               d. None of them.

    Part II: Structured questions 
         
    6. Write the missing word or words on the space before each number.
    The term ………………….. is often used to refer to X-ray CT. 
              e. Gastroscopy is a procedure that enables your surgeon to examine the lining of the ………….
               f. The most sensitive imaging test available for the diagnosis of acute cerebral infarction is …………... 
               g. Array of …………………. to transform the flashes into amplified electrical pulses inside the body.
                h. Transducers used are different depending on ………. of a patient, one has 5 MHz and other 3.5 MHz.
                 i. Hydrogen nuclei (also called protons) behave as small …………… that align themselves parallel to the field.
                 j. In ………………….. there are appearance three words: nuclear, magnetic and resonance.
                 k. Examination can be ………………… is one of the disadvantages of MRI. 
              7.  Answer by True if it is True and by False if it is False
                  a. The use of gamma radiation to form images following the injection of various radiopharmaceuticals is known as Scintigraphy. 
                  b. This decision to scan or not to scan a normal pregnancy must be made only by the photographer. There are universally accepted guidelines at present.
                  c. Tissue in the body absorbs and scatters ultrasound in the same ways. Lower frequencies are more rapidly absorbed (attenuated) than higher frequencies.
                  d. Upper endoscopy uses light and camera to view the esophagus, stomach, and upper part of the small intestine.
                  e. Ultrasound is both generated and detected through high frequency oscillations in piezoelectric crystals so there is ionizing radiation exposure associated with ultrasound imaging.

               8. Compare endoscopy imaging and radionuclide imaging
               9. What are the advantages of MRI in clinical practice?
             10. Is ultrasound safe? explain.
             11. What areas of the body can be imaged by ultrasound?
             12. Why is ultrasound used in pregnancy?
             13. Explain the advantage of CT scan 
            14. In mammography exams, is the breast compression necessary?Why 

    Essay question Historically, MRI began in the central nervous system, but it is now extended to all regions of the human body. The excellent resolution and contrast available in any chosen plane in the body, makes the MRI an invaluable diagnostic tool with which to study body structure, function and chemistry, as well as disease. Discuss the application of MRI.
    Unit 11: EFFECTS OF LASER.UNIT 13: RADIATIONS AND MEDICINE